Model Number 186172680
Medical Device Problem Codes Break (1069); Material Twisted/Bent (2981)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, the surgeon was performing a lengthening on a patient and the two extended connectors implanted.The expedium 4.5 set screw from the connector stripped and broke inside the connector.There was a twenty minute surgical delay.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: a review of the receiving inspection (ri) for exp 4.5 ti ext tand conn 80 was conducted identifying that lot number wa3569 was released in a single batch.Batch1: lot qty of (b)(4) units were released on march 3, 2017 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that both ends of the exp 4.5 ti ext tand conn 80 appear to be stripped.No evidence of breakage was observed.No dimensional inspection was performed due to post-manufacturing damage.Complaint relevant dimensions cannot be taken.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces as mentioned in the event description.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the exp 4.5 ti ext tand conn 80 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM SPINE SYSTEM EXTENDED TANDEM CONNECTOR 4.5 - 4.5 X 80MM
Common Device Name
GROWING ROD SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13828073
Report Number1526439-2022-00423
Device Sequence Number16681803
Product Code PGM
UDI-Device Identifier10705034109145
UDI-Public(01)10705034109145
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K141509
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186172680
Device Catalogue Number186172680
Device Lot NumberWA3569
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer03/21/2022
Initial Report FDA Received Date03/19/2022
Supplement Report FDA Received Date04/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/03/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXP 4.5 TI EXT TAND CONN 40
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-