| Catalog Number |
09P31-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, abbott point of care was contacted by a customer regarding i-stat chem8+ cartridges that yielded suspected discrepant results on a 50 year old female patient.The discrepant results occured at 15:03 with sample a.Patient was suspected to have low hemoglobin.The hemoglobin results varied.There was no additional patient information available at the time of this report.Return product is not available for investigation.(b)(6).There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.It is suspected that the results for 15:03, sample a was a hemolyzed sample.However, this could not be confirmed at the time of this report.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 04-apr-2022.A review of the device history record confirmed the lot passed finished goods release criteria.Retained cartridge testing for the lot met the acceptance criteria found in q04.01.003 rev.Ag, appendix 1- product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.
|
| |
|
Search Alerts/Recalls
|