MAUDE Adverse Event Report: RESPIRONICS, INC. REMSTAR PRO C-FLEX+; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. REMSTAR PRO C-FLEX+; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number 461CA
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
 
Event or Problem Description
The manufacturer received information alleging an issue related to the cpap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
The manufacturer previously reported incorrect information alleging an issue related to the cpap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.Section b5 should have been reported as: a cpap was returned for service with an allegation of "service required, mask filled with smoke".The complaint also indicates the devices were making a "humming/shining sound".There was no patient harm or injury reported.The device was evaluated by the manufacturer and performed an external and internal inspection of the received device and found no evidence of thermal damages.There was evidence of dirt/dust contamination to the air inlet filters (air inlet path of the remstar) and also to the remstar sd card slot.This contamination was later found throughout the remstar air path.The blower assembly was disassemble and evidence of dirt/dust contamination and a dark-colored oil-like substance coating the internal surfaces of the remstar blower assembly was observed.The impeller assembly shows evidence of having fractured likely due to the impeller becoming unbalanced from the build-up of contaminants.The contaminants would also likely result in the degradation of the impeller assembly.There are indications of the impeller assembly having rubbed the blower assembly, likely resulting in the noted complaints.Based on the investigation, the manufacturer does confirm the complaints associated with the returned device.The complaints are found to be due to a failure of the blower impeller in the remstar as a result of improper care of the remstar by the user of the device and is not a result of a failure of the remstar during normal usage.The device was being used in an environment containing the dirt/dust contaminants and a dark-colored oil-like contaminants.These contaminants entered the remstar airpath leading to a failure of the remstar blower assembly.Section e.1 was incorrectly reported in previous report, which is now updated in this follow-up report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR PRO C-FLEX+
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13830074
Report Number2518422-2022-10995
Device Sequence Number991419
Product Code BZD
UDI-Device Identifier00606959030138
UDI-Public00606959030138
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number461CA
Device Catalogue Number461CA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2016
Supplement Date Received by Manufacturer11/22/2016
Initial Report FDA Received Date03/20/2022
Supplement Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Patient SexUnknown
-
-