Model Number 04.038.400S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional narrative: additional product code: ktt initial reporter is a synthes sales representative the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes (b)(6) reports an event as follows: it was reported that on (b)(6) , 2021, the patient underwent an open reduction internal fixation surgery for trochanteric fracture of femur.After surgery, in (b)(6), 2022, the patient had a rear-end accident while driving.A week after the accident, the patient visited the hospital with a complaint of pain and was examined.As a result, a cut-out was found.On (b)(6), 2022, the revision surgery via tha will be performed.The surgeon commented that this cut-out was likely unrelated to the products and caused by the traffic accident and walking after the accident.No further information is available.This report is for a helical blade.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H4 device history: manufacturing location: elmira / packaged, sterilized and released by: monument.Manufacturing date: 02-dec-2019.Expiration date: 01-nov-2029.Part number: 04.038.400s, tfna fenestrated helical blade 100mm -sterile.Lot number: 28p2116 (sterile).Lot quantity: (b)(4).Note: helical blade was manufactured by elmira; lot number 25p7956.Parts were packaged, sterilized and released by monument.Production order traveler met all inspection acceptance criteria.Packaging label log (pll) was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn 16905 supplied by sterigenics was reviewed and determined to be conforming.This lot met all visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.This lot met all visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFNA FENESTRATED HELICAL BLADE 100MM - STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
ELMIRA
35 airport road
horseheads NY 14845
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13830583
Report Number8030965-2022-01759
Device Sequence Number16391543
Product Code HSB
UDI-Device Identifier10886982099297
UDI-Public(01)10886982099297
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.038.400S
Device Catalogue Number04.038.400S
Device Lot Number28P2116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date03/20/2022
Supplement Report FDA Received Date05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/02/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TFNA FEM NAIL Ø11 LE 125° L235 TIMO15; UNK - END CAPS: TFNA; UNK - SCREWS: NAIL DISTAL LOCKING
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-