LEMAITRE VASCULAR, INC ANASTOCLIP GC CLOSURE SYSTEM; CLIP, IMPLANTABLE Back to Search Results
Model Number 4008-06
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
Clip will fall out or clip will be partially crimped after activating the applier.Five or six clips will apply perfectly and then a series of 2-3 that fall out uncrimped or do not close completely after squeezing the applier.
 
Additional Manufacturer Narrative
We have not received the complaint device for evaluation.Hence, we could not conclusively determine the root cause of the incident.Please note that during the in-process inspection, the manufacturing operator visually inspects a sample of the gc clips by checking the tips of the clips are either touching or crossed over to be accepted.A batch review of the affected lot could not be performed since the lot number was not provided to us.We have reviewed our complaint history records for the last 5 years.We have not received any other complaints of a similar nature for anastoclip gc-medium.Hence, we consider this to be an isolated incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANASTOCLIP GC CLOSURE SYSTEM
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
LEMAITRE VASCULAR, INC
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer Contact
pragya thikey
63 second ave
burlington, MA 01803
7812212266
MDR Report Key13830813
Report Number1220948-2022-00019
Device Sequence Number1401271
Product Code MJN
UDI-Device Identifier00840663109579
UDI-Public00840663109579
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K173323
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4008-06
Device Catalogue Number4008-06
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2022
Initial Report FDA Received Date03/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1