• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL INC. 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL INC. 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100B
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/18/2022
Type of Reportable Event Death
Additional Manufacturer Narrative
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.However, a biomed staff who is trained in 3100 device ordered new transducer and will be doing repair of vent.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
It was reported to vyaire medical that the delta p on the 3100 b ventilator was fluctuating between 90cmh2o and 120cmh2o when set to 110cmh2o.The patient was in stable condition the whole time but expired the next day.The customer did not think it was due to the ventilator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Common Device Name
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL INC.
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key13831593
Report Number2021710-2022-15722
Device Sequence Number1404113
Product Code LSZ
UDI-Device Identifier10846446003543
UDI-Public(01)10846446003543(11)20111229
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P890057
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3100B
Device Catalogue Number770155
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2022
Initial Report FDA Received Date03/20/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/29/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-