| Model Number |
TMICL 13.2 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Optical Problem (3001)
|
| Health Effect - Clinical Codes |
Halo (2227); Discomfort (2330); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Weight-race: unk.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 13.2mm tmicl 13.2 implantable collamer lens of -5.50/1.0/070 (sphere/cylinder/axis) was implanted into the patients right eye on (b)(6) 2022.On (b)(6) 2022 the lens was removed and a replacement lens was implanted of shorter length lens due to excessive vault.The problem was resolved.Cause reported as patient related factor.
|
| |
|
Additional Manufacturer Narrative
|
|
B5- patient experienced discomfort and halos (residual refractive error)- should have been included on initial mdr.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3-device evaluation: lens returned dry in wrapped cloth with debris on product.Visual inspection found haptic torn and bent.Debris throughout the lens and one haptic is folded in.Claim# (b)(4).
|
| |
|