• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 SOFT 8MM X 30CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 SOFT 8MM X 30CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035478300
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during the procedure, the subject coil was stretched at the distal part upon deployment, and it could not be advance or retrieved inside the microcatheter.The physician applied force to remove the coil (subject device) but the main coil was fractured and was left inside the anatomy.A snare was used to retrieve the fragment but removal was not successful.There was a 40 mins delay during the procedure to remove the coil.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
Due to the automated mes (manufacturing execution system) system, there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection, as well as a functional evaluation, could not be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.It was reported that during the procedure coil stretched upon attempted deployment into aneurysm.Microcatheter and coil would not advance forward or backward.Physician was forced to remove microcatheter and leave coil tail in patient.Additional information provided by the customer indicated that there was surgical delay of 40 minutes.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable will be assigned to this complaint.
 
Event or Problem Description
It was reported that during the procedure, the subject coil was stretched at the distal part upon deployment, and it could not be advance or retrieved inside the microcatheter.The physician applied force to remove the coil (subject device) but the main coil was fractured and was left inside the anatomy.A snare was used to retrieve the fragment but removal was not successful.There was a 40 mins delay during the procedure to remove the coil.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 SOFT 8MM X 30CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key13831778
Report Number3008881809-2022-00126
Device Sequence Number991468
Product Code HCG
UDI-Device Identifier04546540676603
UDI-Public04546540676603
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2024
Device Model NumberM0035478300
Device Catalogue NumberM0035478300
Device Lot Number22978154
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date03/20/2022
Supplement Report FDA Received Date04/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR SL-10 MICROCATHETER (STRYKER)
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age56 YR
Patient SexFemale
-
-