• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO PGLA HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO PGLA HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number PV-20-50-HELIX
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that there was migration after deployment.The patient was taken to an intra-abdominal vein occlusion procedure.The coils were placed in the right hypogastric vein and the coil migrated and was located in the right pulmonary artery superior lobal branch.The patient was reprogrammed for removal of the foreign body coils and they were successfully removed.The cause of the migration was not determined.The coil ended up being placed in the correct location.The device was prepared as indicated in the ifu.No symptoms or complications were associated with the event.This event did lead to surgical intervention and a hospitalization for reintervention.The patient had normal blood flow and minimal vessel tortuosity.
 
Event or Problem Description
Additional information received reported that the patient was being treated for occlusion of intra-abdominal veins via endovascular approach.Embolization is performed on: left gonadal with concerto type coils 20x50 1 and 16x40 (1) left hypogastric with 20x50 nester type coils (#2) and right hypogastric with 20x50 concerto type coils (#2).Guideand catheter are removed, infusion of ssn by introducer is started, recovery is passed.No embolization of foreign bodies is visualized.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCERTO PGLA HLX
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key13834919
Report Number2029214-2022-00448
Device Sequence Number11149298
Product Code KRD
UDI-Device Identifier00847536033689
UDI-Public00847536033689
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K090046
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/07/2024
Device Model NumberPV-20-50-HELIX
Device Catalogue NumberPV-20-50-HELIX
Device Lot NumberB187429
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date03/20/2022
Supplement Report FDA Received Date04/14/2022
Date Device Manufactured04/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexFemale
-
-