| Model Number |
RONYX20018UX |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/03/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
One resolute onyx coronary drug eluting stent was attempted to be used to treat a moderately tortuous, severely calcified lesion with 99% stenosis in the mid left anterior descending (lad) artery.The device was inspected prior to use with no issues noted.Negative prep was not performed.The lesion was pre-dilated.The device did pass through a previously deployed stent.Resistance was encountered when advancing the device and excessive force was used during delivery.It was reported that stent dislodgement occurred during delivery to the lesion.The physician stated that they think the stent hooked on the previously deployed stent and did not blame the resolute onyx stent.The dislodged stent was removed by passing a small balloon distal to the stent, inflating the balloon minimally and pulling the stent back into the guide extension catheter and the guide catheter.No further patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|