| Model Number |
C111 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer replaced a sensor and this resolved reagent bottle liquid level detection errors the customer received.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for two patient samples tested with crej2 creatinine jaffé gen.2 (compensated) on a cobas c 111.The sample initially resulted in a creatinine value of 4.6 mg/dl and this value was reported outside of the laboratory to the clinician.The clinician questioned the result.On 02-mar-2022, the reagent bottle had 8 tests remaining and was replaced.The sample was then repeated on 02-mar-2022, resulting in a value of 1.0 mg/dl.The second sample initially resulted in a creatinine value of 5.8 mg/dl on 10-mar-2022 and repeated as 0.9 mg/dl on 10-mar-2022.No incorrect results were reported outside of the laboratory for this sample.The creatinine reagent lot number was 57929101, with an expiration date of 30-jun-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon review of the error log files from the instrument, multiple system reagent level detection messages were observed.The field service engineer also observed a sample liquid level detection error.An issue with level detection signal stability was identified.The engineer replaced a transfer head assembly and after this intervention, the instrument performed within specifications.The investigation determined the service actions resolved the issue.
|
| |
|