| Catalog Number |
08057958190 |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of a questionable low ldhi2 lactate dehydrogenase ver.2 result from cobas pro c 503 analytical unit serial number (b)(4).The initial result was 167 u/l.The repeat result at another site was 409 u/l.On (b)(6) 2022, the repeat result using reagent lot number 618368 was 399 u/l.Using a new pack from reagent lot 589772, the result was 409 u/l.It was requested but not known if any questionable result was reported outside of the laboratory.The repeat results were believed to be correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The calibration data provided was acceptable.The issue was resolved by the customer using a new reagent cassette from the same lot.The investigation determined the issue was consistent with contamination or denaturation of a single reagent cassette.The cause of the contamination or denaturation could not be determined with the information available.
|
| |
|