ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION KIT: 19 GA; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Model Number IPN045606
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
Ran into this issue today on l & d.Epidural catheter was sealed at the proximal end.Additional information: the catheter had already been inserted into the patient.Because the defect was the end distal to the patient it was not discovered until the tuochy needle was removed and something caught as the catheter passed through the lumen.I ended up taking sterile scissors and cutting off the sealed portion which allowed the catheter to function as normal.
 
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
Ran into this issue today on l & d.Epidural catheter was sealed at the proximal end.Additional information: the catheter had already been inserted into the patient.Because the defect was the end distal to the patient it was not discovered until the tuochy needle was removed and something caught as the catheter passed through the lumen.I ended up taking sterile scissors and cutting off the sealed portion which allowed the catheter to function as normal.
 
Additional Manufacturer Narrative
(b)(4).A device history record review was performed on the epidural catheter with no relevant findings.A corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed on the epidural catheter with no evidence to indicate a manufacturing related issue.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without a sample.No further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT: 19 GA
Common Device Name
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
312 commerce place
asheboro NC 27203
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key13839474
Report Number1036844-2022-00019
Device Sequence Number910987
Product Code BSO
UDI-Device Identifier40801902005128
UDI-Public40801902005128
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K140110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model NumberIPN045606
Device Catalogue NumberAK-05501
Device Lot Number23F21J0082
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer05/02/2022
Initial Report FDA Received Date03/21/2022
Supplement Report FDA Received Date05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.
Patient SexUnknown
-
-