ARTHREX, INC. 2.4MM BIO-SUTURETAK DISPOSABLES KIT; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
|
Back to Search Results |
|
| Model Number |
2.4MM BIO-SUTURETAK DISPOSABLES KIT |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
During returned device evaluation, a reportable malfunction was discovered.Complaint confirmed.One unpackaged ar-1934-24ds was received for investigation.Visual inspection identified that the drill bit had seized inside the guide.Using digital micrometer id: 224, it was revealed that approximately 9.082mm of the drill bit's distal tip had fragmented.Additionally, the proximal end of the drill bit had also broken off.The cause remains undetermined, although a probable cause can be attributed to prying/leveraging the drill within the guide, leading to friction between the components.
|
| |
|
Event or Problem Description
|
|
On 01/06/2022 it was reported by a arthrex representative via sems that an ar-1934-24ds, the drill bit breakage stacked in the drill guide, the tip of the drill was damaged.The case was completed using the new product in the same lot.No patient harms.Additional information 1/13/2022: on 01/13/2022 it was reported by a arthrex representative via sems that an ar-1934-24ds, during a procedure the drill but was stuck inside the drill guide.The drill tip was damaged during drilling, no debris came off the device.The case was competed using a new ar-1934-24ds.During returned device evaluation, a reportable malfunction was discovered.
|
| |
|