• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; BLOOD GLUCOSE TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; BLOOD GLUCOSE TEST STRIPS Back to Search Results
Model Number 7278
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record was reviewed for the suspected contour next test strips and no manufacturing anomalies were found.The dimensions of the carton makes it difficult for a vial cap to open after packaging when the box is sealed.The patient/family was the initial reporter , so personal information was not entered.No information was captured in as the customer's age and weight were not provided.
 
Event or Problem Description
The customer reported that he opened a box of the contour next test strips and the bottle was already open.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement test strips were sent to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT
Common Device Name
BLOOD GLUCOSE TEST STRIPS
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PHC CORPORATION (REGISTRATION NUMBER: 1000134386)
2131-1 minamigata
toon ehime, 791-0 395
JA   791-0395
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key13839659
Report Number1810909-2022-00048
Device Sequence Number13839428
Product Code NBW
UDI-Device Identifier10301937278706
UDI-Public10301937278706
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111268
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/31/2023
Device Model Number7278
Device Catalogue Number7278
Device Lot Number1MPEH41
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/02/2022
Initial Report FDA Received Date03/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-