• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP AVAPS; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP AVAPS; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Model Number FR1165X
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
 
Event or Problem Description
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP AVAPS
Common Device Name
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13839684
Report Number2518422-2022-10384
Device Sequence Number1401664
Product Code MNS
UDI-Device Identifier00606959039087
UDI-Public00606959039087
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K102465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberFR1165X
Device Catalogue NumberFR1165X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2017
Initial Report FDA Received Date03/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/17/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Patient SexUnknown
-
-