| Model Number |
PS11.5-65SJ250 |
| Medical Device Problem Code |
Material Discolored (1170)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
Information was received that during explant the nail was found to have corrosion on the telescopic part.The nail functioned as intended and no patient adverse event was reported.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection of the returned nail revealed that the male/female junction of the nail (telescopic part) had discoloration, which confirmed crevice corrosion as the failure mode.The crevice corrosion failure is a previously known issue and has been investigated under a capa.Review of the device history records indicated the nail was visually inspected and met all quality inspections prior to release.
|
| |
|