(b)(4).Batch #: p9415y.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.The device was returned with no apparent damage.The device was connected to a test hand piece and a gen11.During functional testing, it was noted that the max hand activation button was nonfunctional.However, the device did activate when tested with the foot switch.No alert screen was displayed as reported.The device was disassembled to inspect the internal components.Corrosion was found at the max hand activation dome.Due to the moisture discovered on the instrument which is evidence of possible reuse, we are unable to determine how this condition impacted the performance of the device and therefore cannot conclude a root cause.Our manufacturing, sterilization, packaging, and shipment processes do not introduce corrosion / moisture to the device.As part of ethicon endo surgery quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|