MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problems No Audible Alarm (1019); Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/26/2021
Event Type  malfunction  
Event Description
Information was received from multiple sources (manufacturer representative, healthcare provider, consumer) regarding a patient who was receiving intrathecal dilaudid (2,000 mcg/ml at 306 mcg/day) via an implantable pump for spinal pain indications.It was reported that multiple motor stalls were discovered when the logs were interrogated occurring between (b)(6) 2021 and (b)(6) 2022.The patient reported that she did not recall hearing any alarms and all stalls recovered in less than 2 hours.She also has not reported any signs of withdrawal or change in therapy.The cause of the motor stalls was unknown. at this time the pump has recovered and no other troubleshooting has been performed.No actions/interventions were taken at this time.It was asked but unknown if surgical intervention was planned.The issue was resolved at the time of this report and the patient's status was "alive- no injury".The patient's weight and medical history were asked and would not be made available.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13840559
MDR Text Key287550491
Report Number3004209178-2022-03543
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169508149
UDI-Public00643169508149
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2017
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/16/2022
Date Device Manufactured06/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0497-2013
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
-
-