ARTHREX, INC. KEEL PUNCH, SIZE 4-7, IBALANCE TKA; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
|
Back to Search Results |
|
| Model Number |
KEEL PUNCH, SIZE 4-7, IBALANCE TKA |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2021, it was reported by a facility representative via sems that an ar-623-t6 tibial prep tray and an ar-623-6 keel punch are broken.This was discovered during a procedure on (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed, the dovetail feature is partially broken.Edges of the device are dented and worn.The cause of the event is undetermined, however a likely cause is user-applied mechanical forces.
|
| |
|
Search Alerts/Recalls
|
|
|