|
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and suture was used.During the procedure, when tying a knot with an instrument during interrupted suturing, the suture was broken at the 1st ligation.After that, three sutures were used, but the same issue occurred.No adverse patient consequences were reported.Additional information was requested.
|
|
Product complaint # (b)(4).Date sent to the fda: 4/19/2022.Additional information: d4, d9, h3, h4, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: visual analysis of the returned product revealed that one empty opened foil, an empty labeled winding former, and a needle-suture piece product code z493g were received to ethicon inc for evaluation.Upon visual inspection of the returned sample, body fluids, wavy, damage at the surface, and split were observed on the suture.After further evaluation crimping marks were observed on the surface suture possibly from a surgical instrument.The product code z493g contains absorbable suture, as the sample was received open, the time of exposure to the environment could not be determined and a functional test cannot be performed.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|