• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO PREMIUM SURGICLIP; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO PREMIUM SURGICLIP; CLIP, IMPLANTABLE Back to Search Results
Model Number 134031
Medical Device Problem Codes Fail-Safe Problem (2936); Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the instrument was fully applied with no clips remaining.The distal end of the channel cover was intact.Microscopic inspection of the jaws revealed acceptable jaw co-planarity.Functional testing noted that the instrument was cycled to complete the firing cycle, the interlock did not engage.The pusher bar was dragging the interlock mechanism backwards.It was reported that the safety interlock deployed prior to all staples or clips firing.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: the interlock was nonfunctional and did not engage.The most likely cause was determined to be manufacturing related.This product is no longer being manufactured.The event had foreseen risks and is included in a data monitoring plan.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during an open procedure, on the firing step, the surgeon was unable to squeeze the handle, the clip applier blocked before loading.Another device was opened to resolve the issue.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREMIUM SURGICLIP
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key13843759
Report Number2647580-2022-00931
Device Sequence Number4132306
Product Code FZP
UDI-Device Identifier20884521057521
UDI-Public20884521057521
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K143644
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model Number134031
Device Catalogue Number134031
Device Lot NumberP0A1215Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2022
Initial Report FDA Received Date03/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-