| Model Number |
IPN040971 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "inflation defect found during operation use".No patient injury or harm reported.Patient condition unknown at time of report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.Functional testing was then performed and the cuff and pilot balloon were inflated to 25mbar using a calibrated endotest meter.The cuff could not stay inflated.The sample was then immersed in water for leak testing.Air bubbles were observed coming from the pilot balloon indicating there was a leak.A device history record review was performed and no relevant findings were identified.The complaint of leakage was confirmed; however, a root cause for the issue could not be determined.The manufacturing site reports there is a 100% water leak test on the pilot balloon and 100% leak testing on the side arm fitting during the manufacturing process whereby any leakage would be detected prior to release from the manufacturing facility.
|
| |
|
Event or Problem Description
|
|
It was reported that "inflation defect found during operation use".No patient injury or harm reported.Patient condition unknown at time of report.
|
| |
|
Search Alerts/Recalls
|