Catalog Number 195-160
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.A supplemental report will be provided after completion.
 
Event or Problem Description
He consumer reported a false positive result with the binaxnow¿ covid-19 antigen self test.She performed an unknown test with her doctor and still got a positive result.The patient was asymptomatic.No additional patient information, including treatment and outcome, was provided.".
 
Additional Manufacturer Narrative
Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 178321 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 195-160/ lot 178321, test base part number 195-430h/ lot 175377.The lot met the required release specifications.A review of the complaints reported as false positive and/or false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 178321 showed that the complaint rate is (b)(4).Abbott diagnostics scarborough, inc.Was unable to determine the exact root cause of the reported issue ; however, it could possibly be related to issues including the self-test user performance or the specific patient sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BINAXNOW COVID-19 SELF-TEST
Common Device Name
LATERAL FLOW IMMUNOASSAY IVD OF COVID-19
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC
10 southgate road
scarborough ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough, ME 04074
MDR Report Key13845512
Report Number1221359-2022-01498
Device Sequence Number4132352
Product Code QKP
UDI-Device Identifier00811877011415
UDI-Public01008118770114151722121710178321
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EUA210264
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/17/2022
Device Catalogue Number195-160
Device Lot Number178321
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2022
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date03/22/2022
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexFemale
-
-