| Model Number |
IPN048573 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "the light of the handles always starts to flicker during intubation, so the handle had to be changed again and again.After use, the handle was checked and the batteries removed.The first battery inserted into product was deformed on behalf of the product, which was not the case before insertion of the battery into the handle".No patient injury or harm reported.Patient condition reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4), the sample was returned and sent to the manufacturing site for investigation.The manufacturing site reports: "we inspected the complaint device and found that the bottom cap of the complaint device was loose and not tightened properly that could be contribute for light flickering of handles and blades.During functional testing of complaint device, it was identified that the bottom cap of the handle was in loose condition and not mounted properly, so firstly we tightened the bottom cap properly and performed the testing of handle and blade, and we have observed that complaint sample was working fine and light from the blade was constant and blowing perfectly.The device history record for lot code 190101 was reviewed and no issue that could have contributed to the reported failure was noted.The device was manufactured according to release specification.The root cause of this complaint is deemed an unintentional user-related error because in the investigation it has been identified that the user had not tightened the bottom cap of the handle properly, therefore the customer had countered the light flickering issue of the handle.".
|
| |
|
Event or Problem Description
|
|
It was reported that "the light of the handles always starts to flicker during intubation, so the handle had to be changed again and again.After use, the handle was checked and the batteries removed.The first battery inserted into product was deformed on behalf of the product, which was not the case before insertion of the battery into the handle".No patient injury or harm reported.Patient condition reported as "fine".
|
| |
|
Search Alerts/Recalls
|