SMITH & NEPHEW, INC. ENDOSCPC CANN.DRL.BIT 4.5 STRL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Model Number |
7207315 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during an arthroscopy procedure, while using the endoscopic cann drill bit it broke.The procedure was successfully completed with no surgical delay using a back-up device.No patient injury or other complications were reported.Results of investigation have concluded that the distal tip of the drill bit is broken from the shaft which makes it a reportable event.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).The sample is under evaluation by the manufacturing site.Preliminary results of investigation found that the reported device was received for evaluation.It was determined the device did not contribute to the reported event.A complaint history review concluded this was an isolated event. a review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution. a visual inspection of the returned device found that it is not in its original packaging.The device is worn from use, and the distal tip of the drill bit is broken from the shaft.There is debris on the device.The complaint was confirmed, and the root cause was associated with unintended use of the device. factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device was received for evaluation.It was determined the device did not contribute to the reported event.A visual inspection of the returned device found that it is not in its original packaging.The device is worn from use, and the distal tip of the drill bit is broken from the shaft.There is debris on the device.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Without the requested information, the reported endoscopic cann drill bit breakage could not be further assessed.It is unknown if the broken fragment was retrieved from the patient.The endoscopic cann drill bit is made of 316l stainless steel.The drill is an externally communicating device, it is neither manufactured nor intended for implantation.It is also not approved for long term internal tissue exposure and long-term implantation data is not available.The patient impact beyond the reported device breakage and possible retained foreign body could not be definitively determined.If retained micro-motion or movement cannot be predicted.There is potential risk for tissue inflammation and/or pain.No further clinical/medical assessment is warranted at this time.Should additional clinical documentation become available in the future, the clinical/medical task may be re-evaluated.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|