Model Number 9014
Medical Device Problem Codes Inflation Problem (1310); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure, the device allegedly failed to inflate.It was further reported that the pinching was noted in the balloon.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation as well as images were provided for review.The investigation of the reported event is currently underway.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one image was reviewed.An angioplasty balloon is seen inflated likely within the distal superficial femoral artery and popliteal artery above the knee.A stent is also seen in part of the segment of artery that is being treated with the balloon.Towards the middle of the balloon, a small area appears as a ¿waist,¿ signifying that it has not inflated properly and is likely twisted.Therefore, based on the image review, the reported failure balloon twist and inflation issue can be confirmed.One lutonix 018 drug coated balloon was returned for evaluation.During the visual evaluation, wrinkles were noted on the returned balloon; no other anomalies were noted on the returned device.On functional testing of the device, the balloon was inflated with the in-house presto inflation device up to 7 atm.The balloon maintained its shape and pressure.On further, the balloon deflated.No other functional testing was performed.All the anomalies noted during the microscopic observation.Therefore, the investigation was confirmed for the reported material twist and inflation issue as wrinkles were noted on the balloon during the visual evaluation of the returned device.Image review also confirms the reported issue.A definitive root cause for the reported material twist and inflation issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 03/2024), g3, h6 (method).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the device allegedly failed to inflate.It was further reported that the pinching was noted in the balloon.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 018 DRUG COATED DILATATION CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
LUTONIX, INC
9409 science center dr
new hope MN 55428
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13854325
Report Number3006513822-2022-00010
Device Sequence Number11113682
Product Code ONU
UDI-Device Identifier00801741177699
UDI-Public(01)00801741177699
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9014
Device Catalogue NumberLX1813061505F
Device Lot NumberGFFN2908
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer04/08/2022
Initial Report FDA Received Date03/22/2022
Supplement Report FDA Received Date04/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-