• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL ARTHSCO,4/140_30_QIK/STRKR HUB; RIGID ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL ARTHSCO,4/140_30_QIK/STRKR HUB; RIGID ENDOSCOPE Back to Search Results
Model Number 242043
Medical Device Problem Codes Display Difficult to Read (1181); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: the complaint device was received at the manufacturing site and evaluated.It was reported that "damaged by wand".Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: tube is bent.Poor image quality damaged optics : objective, window, and optics are damaged.Image is cloudy/foggy.Minor scratches on the unit.The defective parts were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The user error was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event or Problem Description
It was reported by the sales rep that during an unknown shoulder procedure on (b)(6) 2022, it was observed that the endoscope device was damaged by wand.During an in-house engineering evaluation, it was determined that the image from the device was cloudy/foggy.Another like device was used to complete the procedure without delay.There were no adverse patient consequences reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHSCO,4/140_30_QIK/STRKR HUB
Common Device Name
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13858995
Report Number1221934-2022-00893
Device Sequence Number19041596
Product Code HRX
UDI-Device Identifier10886705026234
UDI-Public10886705026234
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K971996
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number242043
Device Catalogue Number242043
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2022
Initial Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-