STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5532-G-409-E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Fibrosis (3167)
Date of Event 02/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
It has been reported "the sae happened on (b)(6) 2022.This patient had functional loss due to arthrofibrosis 8 weeks postoperatively.At (b)(6) 2022 the patient was hospitalized to perform manipulation under anesthesia.No explant or revision surgeries were done or planned.The event is related to the procedure (total knee arthroplasty), but we cannot say yet whether this is also related to this specific implant.".
 
Additional Manufacturer Narrative
Reported event: an event regarding rom/arthrofibrosis involving a triathlon insert was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient underwent a medical procedure to address functional loss and arthrofibrosis of the knee.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.Catalog numbers and lot codes of other devices listed in this report: (b)(6), tritanium bplate triathlon s4, lot ctd58977.(b)(6), triathlon p/a ps beaded #5l, lot l7n7d.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event or Problem Description
It has been reported "the sae happened on (b)(6) 2022.This patient had functional loss due to arthrofibrosis 8 weeks postoperatively.At (b)(6) 2022 the patient was hospitalized to perform manipulation under anesthesia.No explant or revision surgeries were done or planned.The event is related to the procedure (total knee arthroplasty), but we cannot say yet whether this is also related to this specific implant.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key13859413
Report Number0002249697-2022-00428
Device Sequence Number16371365
Product Code MBH
UDI-Device Identifier07613327338355
UDI-Public07613327338355
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K172634
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model Number5532-G-409-E
Device Catalogue Number5532-G-409-E
Device Lot Number3866AK
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age59 YR
Patient SexUnknown
Patient Weight83 KG
-
-