| Model Number |
68-180 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Health care professional reported left side deflation.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: opening curved on radius and brown particles in fill channel.Leak test and microscopic analysis was performed which identified: striated opening on radius and observed void >1 mm in non-textured, valve-to shell-bond area.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a striated opening on radius assessed as surgical damage.Additional, changed, and/or corrected data.
|
| |
|
Event or Problem Description
|
|
Health care professional reported left side deflation.The device has been explanted.
|
| |
|