• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR Back to Search Results
Model Number MM200
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 03/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been made available for evaluation.Should further information of the device become available, a follow-up report will be issued.
 
Event or Problem Description
Consumer alleges while reclining in the chair, the wand where the cord attaches allegedly started shorting out/burning and the consumer allegedly saw sparks and flames.
 
Additional Manufacturer Narrative
The root cause of the failure appears to be a damaged harness, the pcb and all controls of the hand-control had not failed.
 
Event or Problem Description
Consumer alleges while reclining in the chair, the wand where the cord attaches allegedly started shorting out/burning and the consumer allegedly saw sparks and flames.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
ELECTRIC POSITIONING CHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
401 york ave
duryea, PA 18642
8008008586
MDR Report Key13860134
Report Number2530130-2022-00024
Device Sequence Number20006551
Product Code INO
UDI-Device Identifier00606509500081
UDI-Public00606509500081
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K070950
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMM200
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date08/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-