• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. J-VAC RESERVOIR 300 ML; CATHETER TIP SUCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. J-VAC RESERVOIR 300 ML; CATHETER TIP SUCTION Back to Search Results
Model Number 2163
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Attempts to obtain the device however not received.If further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported a patient underwent an abdominoplasty on (b)(6) 2022 and a reservoir was used.The piece of metal that is inside the reservoir broke, stopping the suction.No adverse patient consequences were reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information: d9, h6 component code: g07002 product not confirmed.H3 evaluation: the product was returned for evaluation.Material locking mechanism of reservoir was checked to verify its condition.Sample were evaluated, and during this evaluation it was notice that the latch of plate and its mechanism was in good condition, in addition to, it was able to be activated and de-activated several times with no issues.Therefore, is considered this material factor does not contribute to the reported complaint.Machine welding around the ports and in the perimeter of the bag was inspected and it was in good condition, there was no presence of weak welding or over welding that could cause a leakage.The reservoir was activated while the caps were inserted on the ports expecting that no air will go into the reservoir, then, the bag did not inflate.It means that there is no damage on the film of the reservoir.Therefore, it is considered that machine factor does not contribute to the reported complaint defect.Method reservoir didn't present any damage or tears on the front or the back side.Plate was activated, port closed, and the swelled reservoir was pressed to verify if there could be any visible leak.No leak could be identified.Therefore, is considered this method factor does not contribute to the reported complaint defect.Based on the present analysis and sample evaluation, the reported defect by the customer could not be confirmed or related to manufacturing process and other external causes could have affected the product integrity such as handling, non-proper usage or any other possible contributor out of the manufacturing control.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
J-VAC RESERVOIR 300 ML
Common Device Name
CATHETER TIP SUCTION
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-FLEXTRONICS MEDICAL
1659 gailes blvd
san diego CA 92154
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13860296
Report Number2210968-2022-02017
Device Sequence Number14788734
Product Code JOL
UDI-Device Identifier10705031003323
UDI-Public10705031003323
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
CL II EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model Number2163
Device Catalogue Number2163
Device Lot NumberJU0280
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer03/23/2022
08/25/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date04/19/2022
08/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-