Product complaint # (b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint # (b)(4).Additional information: d9, h6 component code: g07002 product not confirmed.H3 evaluation: the product was returned for evaluation.Material locking mechanism of reservoir was checked to verify its condition.Sample were evaluated, and during this evaluation it was notice that the latch of plate and its mechanism was in good condition, in addition to, it was able to be activated and de-activated several times with no issues.Therefore, is considered this material factor does not contribute to the reported complaint.Machine welding around the ports and in the perimeter of the bag was inspected and it was in good condition, there was no presence of weak welding or over welding that could cause a leakage.The reservoir was activated while the caps were inserted on the ports expecting that no air will go into the reservoir, then, the bag did not inflate.It means that there is no damage on the film of the reservoir.Therefore, it is considered that machine factor does not contribute to the reported complaint defect.Method reservoir didn't present any damage or tears on the front or the back side.Plate was activated, port closed, and the swelled reservoir was pressed to verify if there could be any visible leak.No leak could be identified.Therefore, is considered this method factor does not contribute to the reported complaint defect.Based on the present analysis and sample evaluation, the reported defect by the customer could not be confirmed or related to manufacturing process and other external causes could have affected the product integrity such as handling, non-proper usage or any other possible contributor out of the manufacturing control.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|