Model Number 3CX*FX25RWC
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2022
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during non clinical activity, there was black lint within the housing.No patient involvement.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 22, 2022.Upon further investigation of the reported event, the following information is new and/or changed: h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 10, 11, 3331, 4244, 4307).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 3331 - analysis of production records.Type of investigation #3: 4114 - device not returned.Investigation finding: 4244 - packaging contains unintended material.Investigation conclusions: 4307 - cause traced to component failure.The affected sample was inspected upon receipt to confirm a small piece of foreign material on the vr filter.The foreign material was measured and found to be 0.15mm²; which is within specification.Representative retention sample was inspected and noted to have no loose foreign material within the housing.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25RWC W/ RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key13860344
MDR Text Key287672594
Report Number1124841-2022-00061
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450837
UDI-Public(01)00699753450837
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Model Number3CX*FX25RWC
Device Catalogue NumberN/A
Device Lot NumberZD29
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-