Model Number 384539 |
Device Problems
Crack (1135); Fluid/Blood Leak (1250)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 02/25/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
|
|
Event Description
|
Picc line placed as replacement picc on per form (b)(6) penn hospital #2 hub cracked and leaked at 1755pm on (b)(6) 2022.Per customer approximately 23 hours after insertion.Picc d/c¿ed.Piv placed because patient on tpn, lipids, and antibiotics this is the same patient as agh west penn hospital per form #2.Patient had two picc insertions within approximately 23 hours.
|
|
Manufacturer Narrative
|
A review of the dhr and inspection records was conducted, and no similar concerns were found.One sample was returned for review.Visual inspection of the returned sample confirmed a crack in the hub which could result in leakage.Several complaints have previously been received regarding a cracked hub resulting in leakage, and capa 2021-039 has been initiated to address this issue and is currently under investigation.The capa will identify the specific causes and corrective actions to be taken.Capa 2021-039 has been initiated to address this issue and is currently under investigation.The capa will identify the specific causes and corrective actions to be taken.
|
|
Event Description
|
Picc line placed as replacement picc on per form agh west penn hospital #2 hub cracked and leaked at 1755pm on 3/13/2022.Per customer approximately 23 hours after insertion.Picc d/c¿ed.Piv placed because patient on tpn, lipids, and antibiotics this is the same patient as agh west penn hospital per form #2.Patient had two picc insertions within approximately 23 hours.
|
|