• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM Back to Search Results
Model Number 384539
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem Insufficient Information (4580)
Event Date 02/25/2022
Event Type  malfunction  
Manufacturer Narrative
The sample is indicated as available for return. As of the date of this report, the sample has not been returned. A follow-up report will be provided once the device has been received and reviewed.
 
Event Description
Picc line placed as replacement picc on per form (b)(6) penn hospital #2 hub cracked and leaked at 1755pm on (b)(6) 2022. Per customer approximately 23 hours after insertion. Picc d/c¿ed. Piv placed because patient on tpn, lipids, and antibiotics this is the same patient as agh west penn hospital per form #2. Patient had two picc insertions within approximately 23 hours.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameL-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM
Type of DeviceL-CATH PICC
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key13860404
MDR Text Key295899635
Report Number0001625425-2022-00967
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00886333209934
UDI-Public00886333209934
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091670
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation
Type of Report Initial
Report Date 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number384539
Device Catalogue Number384539
Device Lot Number11403016
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/14/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No

Patient Treatment Data
Date Received: 03/23/2022 Patient Sequence Number: 1
-
-