• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FEM POR PS LT SZ 3; ATTUNE CEMENTLESS IMPLANTS : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FEM POR PS LT SZ 3; ATTUNE CEMENTLESS IMPLANTS : KNEE FEMORAL Back to Search Results
Model Number 1504-11-103
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Necrosis (1971); Pain (1994)
Date of Event 03/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, the patient underwent the tka surgery for oa.About 1 month after the surgery, the patient started to feel pain in the affected area.The alignment seemed to be disordered, but the cause of the pain was unknown, and it was decided to perform an open examination of the joint on (b)(6) 2022.On (b)(6) in the open examination, it was found that the distal femur was sinking.Therefore, the implant was exchanged in the revision surgery.The bone surface appeared to have fragility fracture and osteonecrosis, but the cause of this condition was unknown.Since the patient was female, it may be due to osteoporosis.The surgeon commented that the event was not considered to be implant related as there was no problem with tibia.No further information is available.Doi: (b)(6) 2021; dor: (b)(6) 2022; unknown knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.H10 additional narrative:.
 
Event or Problem Description
Additional information was received indicating that there was no surgical delay and no bone cement used that depuy manufactured.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FEM POR PS LT SZ 3
Common Device Name
ATTUNE CEMENTLESS IMPLANTS : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13860477
Report Number1818910-2022-05179
Device Sequence Number14619125
Product Code JWH
UDI-Device Identifier10603295041887
UDI-Public10603295041887
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1504-11-103
Device Catalogue Number150411103
Device Lot Number9825902
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer04/13/2022
04/20/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date04/14/2022
04/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK KNEE ALL POLY TIBIA ATTUNE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-