| Model Number |
48393002 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an avs navigator bayoneted uplif inserter was 'cured/baked'.The device was not lubricated during the inspection as per surgical technique.Surgery was successfully completed with another 'impactor' and a 30 minute delay.
|
| |
|
Event or Problem Description
|
|
It was reported that an avs navigator bayoneted uplif inserter was 'cured/baked'.The device was not lubricated during the inspection as per surgical technique.Surgery was successfully completed with another 'impactor' and a 30 minute delay.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.Device history records could not be reviewed without either the device or a valid lot number.Complaint history was reviewed and no similar complaints for the reported catalogue number were identified.Per the communication with field, the device was not being lubricated.The label on the device states, "navigator bayonet inserter lubricate regularly." the most likely root cause of the reported event is that the user was not lubricating the device.
|
| |
|