NUVASIVE, INCORPORATED NUVASIVE ACP SYSTEM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Model Number 18008317
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2021 a patient underwent an anterior cervical procedure from c4 to c7.The procedure was completed without incident.On (b)(6) 2022 it was discovered during a post-op follow up x-ray that a construct screw had backed out at the c7 level.The patient was experiencing no adverse effects as a consequence of the issue and no revision is planned.
 
Additional Manufacturer Narrative
No product was received as the involved device currently remains in situ with no plans for revision at this time.No conclusive radiograph could be provided confirming alleged product fault.No bone quality report could be provided.The patient is reported to have followed post-op instructions and no fall or other accident could be reported.The root cause of the event is unknown but could be the results of excessive angulation during screw placement.No additional investigation can be completed.Label review: potential adverse events and complications as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications known to occur.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s)¿" ".Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.These devices are not intended to be used as the sole support for the spine.While proper selection can minimize risks, the size and shape of patient anatomy may present limitations on the size of the chosen implants." ".Post-operative warnings - during the postoperative phase it is of particular importance that the physician keeps the patient well informed of all procedures and treatments.Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE ACP SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key13861574
Report Number2031966-2022-00062
Device Sequence Number14795340
Product Code KWQ
UDI-Device Identifier00887517065179
UDI-Public887517065179
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K201078
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number18008317
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2022
Initial Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-