| Model Number |
468-495 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Fistula (1862); Failure of Implant (1924); Unspecified Infection (1930)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "expander was ruptured" against device.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation:analysis of the returned device identified: opening, crease fold, red particles on the surface of the shell.Leak test was performed and found opening on posterior.A microscopic analysis was performed which identify an opening crease sharp in the posterior side.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a crease sharp opening on posterior due to a fold flaw opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "expander was ruptured" against device.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The events of capsular contracture, infection, and fistula are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Reason for reoperation: capsular contracture, infection, and fistula.
|
| |
|
Event or Problem Description
|
|
Company representative reported "deflation, infection, anxiety, capsular contracture baker grade iv, breast hematoma, subsequent fistula formation to the underlying breast implant, tract formation against device." healthcare professional later additionally reported ¿ cancer and late insect envenomation injury with subsequent cellulitis" which are not related to the device.This record is for the left side.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Fi summary: with the information collected during the investigation, there is enough evidence to support that device serial number (b)(6) was manufactured under controlled conditions in accordance with allergan procedures.Seal strength results were acceptable.Environmental monitoring results were found to be acceptable, and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.The sterilization run 30028051 is not related to any additional complaint infection record reported as of today.During the trend review of all infection complaints for saline breast implants for the period of (b)(6)2020 through (b)(6)2022, were noted outliers in mar 2021, oct 2021 and dec 2021.An additional analysis was performed to determine any pattern or any similarities relation between prs involved.It was found that some primary packaging operators were involved in more than one occasion, however the people were trained in the corresponding procedure.Also, it was found a sterilizer involved in more than one occasion on those processes, also, calibrations and maintenance of the equipment¿s involved in more than one occasion were reviewed and it was determined that they were performed on time and met specifications.In addition, all the records involved in this month were reviewed and no issues were found associated to either event.Given the fact that the cause of the infection cannot be specifically associated to the manufacturing process, and there is not an adverse trend for this type of event no corrective action is deemed necessary at this time.
|
| |
|
Event or Problem Description
|
|
Company representative reported "deflation, infection, anxiety, capsular contracture baker grade iv, breast hematoma, subsequent fistula formation to the underlying breast implant, tract formation against device." healthcare professional later additionally reported ¿ cancer and late insect envenomation injury with subsequent cellulitis" which are not related to the device.This record is for the left side.The device has been explanted.
|
| |
|