Model Number 42856
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Codes Loss of Vision (2139); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event occurred in (b)(6) no products have been returned to medtronic for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a shunt.It was reported that during the latest visits, the doctor noticed that the shunt pressure had been changed comparing to the last check (ex: instead of 1.5 it was 2.5).The patient did not have an mri scan.The shunt has been implanted since (b)(6) 2017.The patient is in danger of losing sight.This was reported outside of a procedure.
 
Event or Problem Description
Additional information was received stating that the reported cause of the event is unknown.Deterioration of vision has occurred in the patient as a result of the issue.The patient was being treated for a intra cranial hypertension with a cyst.The patient consulted with the health care professional with loss of vision.A medtronic programmable shunt had been implanted with mild improvements of their vision.The device was maintained at 1.5 but was noticed initially that the device was always changing after mri acquisition and had to checked and modified.In the last two visits it was noticed that the device was changing without any mri examination.The device was found at a level 2 and had to be modified twice to be at level 1.
 
Additional Manufacturer Narrative
A) patient information received.H2) additional information was added to b5.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received.It was reported that on (b)(6) 2022, the patient had a control with his doctor, and the shunt p ressure changed (from 1.5 to 2.It was noted that the doctor "will replace it in the following days.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA®
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key13861851
Report Number2021898-2022-00074
Device Sequence Number5346417
Product Code JXG
UDI-Device Identifier00613994804631
UDI-Public00613994804631
Combination Product (Y/N)N
Initial Reporter CountryTS
PMA/510(K) Number
K042465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2018
Device Model Number42856
Device Catalogue Number42856
Device Lot NumberE00038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer03/28/2022
03/28/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date04/01/2022
04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/15/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age16 YR
Patient SexFemale
Patient Weight50 KG
-
-