WILSON-COOK MEDICAL INC ACUSNARE POLYPECTOMY SNARE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
|
Back to Search Results |
|
| Model Number |
G22700 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report of kinking at the base of the handles.Both devices returned with the snare fully retracted into the sheath.On both devices the handle was manipulated and the snare heads would open and close with no resistance felt.During a visual inspection of both devices, there was damage/kinks noted on the device sheath at the base of the handles.The continuity from the electrical pins to the snare head was tested with an ohm meter and both devices passed.An additional functional test with the devices was performed by attaching the active cord to the electrical pin.The active cord connected to the device easily and remained securely connected.The devices were connected to a valley lab generator and power was applied.Both snares cut simulated tissue as expected.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The instructions for use states: "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." instructions for use states: "fully retract and extend snare to confirm smooth operation of device." instructions for use states: "inspect active cord.Cord must be free of kinks, bends, breaks, and exposed wires to allow for accurate transfer of current.If an abnormality is noted, do not use active cord." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During a polypectomy, the physician used two (2) cook acusnare polypectomy snare.The snares were being used in a colonoscopy to remove a 5cm pedunculated polyp.When the polypectomy was attempted with the first snare, the snare would not cut through the polyp with cautery.They then switched snares to the second snare from the same lot number and encountered the same issue.They switched cautery machines but that made no difference.They switched to a third snare of a different lot number, and were able to finish the procedure.The proximal end of the catheter on both snares, next to the handle, are crumpled up.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|
|
|