| Model Number |
TRIGL |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable triglyceride results from cobas 8000 cobas c 502 module serial number (b)(4).The customer repeated patient samples because they have a procedure to retest samples if the qc performed afterward was out of range.One example was an initial result of 273 mg/dl and a repeat result of 392 mg/dl.The initial result was reported outside of the laboratory.The repeat result was believed to be correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service engineer found the customer recalibrated the assay, which appeared to resolve the issue.Afterward recalibration, qc results were acceptable.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation determined the customer recalibration resolved the issue.
|
| |
|