| Model Number |
03.045.022 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: without a lot number the device history records review could not be completed.Product was not returned.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, during an unknown procedure, when inserting the proximal static screw, the drill bit was getting caught on the nail posterior to the hole.The screw was finally inserted behind the nail.The proximal dynamic screw was inserted properly with no issues using the long self-retaining power shafts from the ans locking set.There was a surgical delay of twenty (20) minutes.Surgery was completed successfully.No fragments generated.This report is for one (1) drill bit/ calibrated/ 4.2 extra-long this is report 5 of 6 for (b)(4).
|
| |
|
Event or Problem Description
|
|
This is report 5 of 7 for pc (b)(4).
|
| |
|
Search Alerts/Recalls
|