• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Cardiac Perforation (2513)
Event Date 03/15/2022
Event Type  Injury  
Event Description
During a cti line in the right atrium, following an atrial fibrillation pulmonary vein isolation ablation procedure on the left side, a tamponade occurred following a steam pop during the last few ablations.The patient underwent a thoracotomy to repair the tamponade and has recovered after a successful surgery.There were no performance issues with the catheter.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The files provided were unable to be analyzed.The event description indicated ablations were performed with an rf power of 40-50 w with an irrigation flow of 17 ml/min.The tacticath sensor enabled contact force ablation catheter instructions for use (ifu) warns that ¿application of rf energy at a power higher than 30w is associated with a higher likelihood of audible steam pop occurrence.¿ in addition, the ifu recommends a minimum irrigation flow of 30 ml/min during ablations using power settings greater than 30 w; however, due to unknown procedural conditions, we are unable to conclusively determine the cause of the steam pop and tamponade.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13862389
MDR Text Key287696758
Report Number3005334138-2022-00257
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-