ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Device Problem
Patient Device Interaction Problem (4001)
|
Patient Problem
Cardiac Perforation (2513)
|
Event Date 03/15/2022 |
Event Type
Injury
|
Event Description
|
During a cti line in the right atrium, following an atrial fibrillation pulmonary vein isolation ablation procedure on the left side, a tamponade occurred following a steam pop during the last few ablations.The patient underwent a thoracotomy to repair the tamponade and has recovered after a successful surgery.There were no performance issues with the catheter.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.The files provided were unable to be analyzed.The event description indicated ablations were performed with an rf power of 40-50 w with an irrigation flow of 17 ml/min.The tacticath sensor enabled contact force ablation catheter instructions for use (ifu) warns that ¿application of rf energy at a power higher than 30w is associated with a higher likelihood of audible steam pop occurrence.¿ in addition, the ifu recommends a minimum irrigation flow of 30 ml/min during ablations using power settings greater than 30 w; however, due to unknown procedural conditions, we are unable to conclusively determine the cause of the steam pop and tamponade.
|
|