• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA SURGICAL OPTICS, INC. B1PC; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA SURGICAL OPTICS, INC. B1PC; INTRAOCULAR LENS Back to Search Results
Model Number B1PC (+22.50 D)
Medical Device Problem Code Sharp Edges (4013)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 01/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Intra-operative complications.Posterior capsule rupture.Doctor reported that something happened in the prepping of the lens or in the injector that is cutting the lens leading haptic.This is creating a sharp edge that is causing the posterior capsule to tear when the sharp edge on the haptic comes in contact with it.Lens still in the eye via reverse optic capture.Patient has recovered and is fine; patient has 2020 and 2030 uncorrected.
 
Additional Manufacturer Narrative
This initial and final emdr is being submitted to fda for a reportable event that occurred in the usa."posterior capsule rupture" is indicated as a potential adverse event related to iol implantation in hoya ifu covered under the warnings section.Haptic damage is indicated as a potential malfunction related to the iol, as covered under the warnings section of the product's instructions for use (ifu).The product was not returned to the manufacturer.The investigation was conducted, with the methods and results as noted below.No abnormalities were found in production and inspection records of the product.(serial no.: (b)(4); model: b1pc).Based on the available information and our investigation, we believe this event was not caused by our product quality.A review of the most recent complaint trending data indicates that no significant trends have been identified at this time and no capa is required as part of the product evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
B1PC
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
HOYA SURGICAL OPTICS, INC.
15335 fairfield ranch rd.
suite 250
chino hills CA 91709
Manufacturer (Section G)
HOYA MEDICAL SINGAPORE PTE LTD
455a jalan ahmad ibrahim
singapore, singapore 63993 9
SN   639939
Manufacturer Contact
goutham pendyala
15335 fairfield ranch rd.
suite 250
chino hills, CA 91709
9093896317
MDR Report Key13862667
Report Number3006723646-2022-00043
Device Sequence Number5296624
Product Code HQL
UDI-Device Identifier00886158017233
UDI-Public00886158017233
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P080004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberB1PC (+22.50 D)
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2022
Initial Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/31/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-