| Model Number |
1116153 |
| Medical Device Problem Codes |
Circuit Failure (1089); Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
|
| |
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported the manufacturer received information alleging an issue related to a bipap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per (b)(4).A correction is being sent due to the initial allegation stating ventilator inoperative alarm conditions were observed in the device's downloaded event log.The device's main board was replaced to address the issues.The sd card was replaced due to errors accessing the filed on the card.The blower was replaced due to noise.The bottom enclosure was replaced due to physical damage.The air inlet foam was replaced due to damage.Other non-actionable errors were alleged but were not confirmed.Sections d8, d9, g1, g3, h2, h3, and h6 have been updated to reflect additional information.
|
| |
|