• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD ARCHITECT PROLACTIN REAGENT KIT; RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD ARCHITECT PROLACTIN REAGENT KIT; RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN) Back to Search Results
Model Number 7K76-25
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer stated that falsely elevated architect prolactin results were generated for a female patient.The initial result was 120 ng/ml and the repeat result was 124 ng/ml (above the normal range).The expected range for females per the architect prolactin package insert is 5.18 - 26.53 ng/ml.When tested on the roche platform at another hospital, the result was within the normal range (exact value not provided).It is suspected the falsely elevated results are due to an interfering substance in the sample.No adverse impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Additional Manufacturer Narrative
Trending review determined no adverse trend for the issue for the product.Historical complaint review determined there is normal complaint activity for the lot number.Device history record review did not show any potential non-conformances, or deviations.Labeling was reviewed and found to adequately address the issue under review.Accuracy testing was completed using panels which mimic patient samples using an in-house retained kit stored at the recommended storage condition.All specifications were met indicating that the lot is performing acceptably.Based on the investigation, no deficiency for lot number 28258ud01 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT PROLACTIN REAGENT KIT
Common Device Name
RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key13862720
Report Number3005094123-2022-00051
Device Sequence Number14788747
Product Code CFT
UDI-Device Identifier00380740014865
UDI-Public00380740014865
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/03/2022
Device Model Number7K76-25
Device Catalogue Number07K76-25
Device Lot Number28258UD01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I2000SR INST, 03M74-02, ISR62793; ARC I2000SR INST, 03M74-02, ISR62793
Patient SexFemale
-
-