| Model Number |
CAX700T12 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
|
| |
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to the bipap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer is submitting a correction for section b5 of the initial report.Section b5 of the initial report should indicate below: the manufacturer received information alleging an issue related to the bipap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.The device was returned to the manufacturer for investigation.During the evaluation of the device, the manufacturer visually inspected the device and found evidence of blower foam degradation, foam particles, and water ingress.Device log files were successfully downloaded and reviewed.The device was repaired and passed final testing.
|
| |
|