• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LAKE REGION MEDICAL LIMITED BENTSON TYPE PTFE COATED GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LAKE REGION MEDICAL LIMITED BENTSON TYPE PTFE COATED GUIDEWIRE Back to Search Results
Model Number M0066201250
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No device was provided for analysis.As reported by customer: "during the case a small piece of the end of the wire broke off, broken piece was retrieved''.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.At this time it is not possible to assign a definitive root cause for the event as reported.Based on the information provided by the supporting documentation, "excessive force used" and/or "improper use" may have impacted on the event as reported.
 
Event or Problem Description
Event description: please provide a brief summary of the event that includes any additional details that you have not already provided: during the case a small piece of the end of the wire broke off, broken piece was retrieved.What troubleshooting steps took place? what troubleshooting steps, if any, resolved the issue?: retrieved the broken piece.What is the next course of action?: opened a new wire.Patient present at time of event?: yes.Patient complications: no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENTSON TYPE PTFE COATED GUIDEWIRE
Common Device Name
GUIDEWIRE
Manufacturer (Section D)
LAKE REGION MEDICAL LIMITED
butlersland
new ross
co. wexford, Y34 K 825
EI  Y34 K825
Manufacturer (Section G)
LAKE REGION MEDICAL LIMITED
butlersland
new ross
co. wexford, Y34 K 825
EI   Y34 K825
Manufacturer Contact
suzanne foley
butlersland
new ross
co. wexford, Y34 K-825
EI   Y34 K825
MDR Report Key13865272
Report Number9681477-2022-00003
Device Sequence Number5272465
Product Code DQX
UDI-Device Identifier08714729056669
UDI-Public(01)08714729056669(17)250930(10)6582044
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K935170
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model NumberM0066201250
Device Catalogue Number620-125
Device Lot Number6582044
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2022
Initial Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-