• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX; LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX; LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) Back to Search Results
Model Number CB030
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
Procedure performed: laprascopic hernia.Event description: the device was being used when performing alf deacon - laprascopic hernia.Scissors failed to cut tissue.It did not jam, the jaws closed but did not cut the tissue.Surgeon then opened another pair of epix laprascopic scissors and completed the surgery with those.There were no changes in patient health resulting from the event.Device replaced with same model scissors and successfully completed surgery intervention: device replaced with same model scissors and successfully completed surgery.Patient status: no complications arose from the event.
 
Additional Manufacturer Narrative
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Testing was performed on the event unit, which confirmed that the returned unit was unable to cut consistently across the entire length of the blade.Engineering observed a small gap between the tips of the blades.Based on the condition of the event unit and description of the event, the scissors were unable to cut due to the gap that were observed between the tips of the blades on the returned unit.  however, the exact root cause of the gap could not be determined based on the evaluation of the returned unit.Applied medical continuously seeks to improve the form, function, and ease of use of its products.  as part of this process, applied medical is currently researching possible enhancements intended to further minimize the potential for this type of event to occur.Correction: sections d4 and h4 are being updated as the lot number was confirmed upon evaluation of the returned unit.
 
Event or Problem Description
Procedure performed: laprascopic hernia.Event description: the device was being used when performing alf deacon - laprascopic hernia.Scissors failed to cut tissue.It did not jam, the jaws closed but did not cut the tissue.Surgeon then opened another pair of epix laprascopic scissors and completed the surgery with those.There were no changes in patient health resulting from the event.Device replaced with same model scissors and successfully completed surgery intervention: device replaced with same model scissors and successfully completed surgery.Patient status: no complications arose from the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX
Common Device Name
LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key13865282
Report Number2027111-2022-00487
Device Sequence Number20151646
Product Code HET
UDI-Device Identifier00607915110123
UDI-Public(01)00607915110123(17)240914(30)01(10)1429335
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K062169
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/14/2024
Device Model NumberCB030
Device Catalogue Number101419701
Device Lot Number1429335
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer02/22/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-