• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETABULAR CUP HAP SIZE 52/60; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETABULAR CUP HAP SIZE 52/60; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74122160
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Metal Related Pathology (4530)
Date of Event 02/12/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Us legal.Bilateral patient.It was reported that the patient underwent a revision surgery of the left hip on (b)(6) 2019 due to metallosis, chronic and increasing pain, failure of the device.The primary left bhr surgery was performed on (b)(6) 2008.The patient underwent another revision surgery of the right hip on (b)(6) 2020.The patient current status is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H6: it was reported that revision surgery was performed on the patients left hip.As of today the devices used in treatment have been requested but have not become available.Without definitive lot numbers, a complete complaint history review cannot be performed for the devices involved.As no device batch numbers were provided for investigation, a manufacturing record review, device labelling and ifu review could not be performed.A risk management review was performed for the bhr head and cup.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.A review of prior escalation actions could not be completed due to insufficient information.The available medical information was reviewed.Based on the reported symptoms it cannot be concluded that the events/clinical reactions of metallosis, chronic and increasing pain, failure of the device were associated with a mal-performance of the implant.The patient impact cannot be determined at this time.Without additional information we cannot further investigate or confirm the reported complaint.The investigation remains inconclusive, and a definitive root cause cannot be determined.Additionally, specific factors known to contribute to the alleged fault cannot be provided due to the insufficient information.If the products or additional information become available in the future, this case will be reopened.Based on this investigation, the need for corrective or preventative action is not indicated.
 
Event or Problem Description
Bilateral patient.It was reported that, after an initial left bhr on (b)(6) 2008, plaintiff underwent a revision surgery on (b)(6) 2009, where a mild effusion was noted intra-articularly; no clear metallosis, only a mild dark staining of intra-articular capsule and soft tissues.Acetabular component was well fixed.(right hip cover under (b)(4) plaintiff outcome is unknown.
 
Additional Manufacturer Narrative
G4, h3, h6: it was reported that a revision surgery was performed on the patients left hip.Mild effusion and mild dark staining were observed intraoperatively.The acetabular component was well fixed.As of today, the implanted device used in treatment are not accessible for testing.Without a definitive batch number, a complete review of the historical complaints data cannot be performed for the acetabular cup.A review of historical complaints data was performed using the part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for the part number and the reported/related failure mode, however, these complaints are from the same patient.This will continue to be monitored.As no device batch number was provided for investigation, manufacturing record review could not be performed.If more information is received, this investigation will be reopened.The review of the current ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the product and similar complaint events was performed.Following the review, no prior applicable escalation actions were identified.The available medical documents were reviewed.The revision report indicates that mild effusion was noted in the articulation and a mild dark staining was noted in the articulation and soft tissues.The operative report states that, in the primary surgery, the acetabular cup was implanted in about 50 degrees of abduction and about 20 degrees of anteversion.It should be noted the bhr surgical technique indicates that the acetabular component must be impacted with 15-20 degrees of anteversion and 40-45 degrees inclination angle.With the information provided, the clinical root cause of the reported pain, mild effusion, and dark staining of the intra-articular capsule and soft tissues cannot be confirmed.It cannot be concluded the reported clinical reactions were associated with a mal performance of the implant or implant failure.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, malalignment of the components and/or soft tissue imbalance may cause excessive wear and early implant failure.Should the device or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETABULAR CUP HAP SIZE 52/60
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13865299
Report Number3005975929-2022-00076
Device Sequence Number17316485
Product Code NXT
UDI-Device Identifier03596010552327
UDI-Public03596010552327
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2019
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74122160
Device Catalogue Number74122160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer04/27/2022
10/07/2022
11/11/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date05/05/2022
10/17/2022
11/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization; Required Intervention;
Patient SexMale
-
-